Rating | 150 deg C, 75 Vac, Cable flame. |
|
Conductor | 36 AWG - 20 AWG. |
|
Insulation | FEP, 7 mils minimum average thickness, 6 mils minimum thickness at any point. |
|
Assembly | Two or more insulated singles optionally twisted around a flouropolymer filler or breather tube. Fillers may be used to form round cable. |
|
Inner jacket | Optional, Silicone rubber, 20 mils min average, 24 mils min at any point.
. |
|
Shield | Optional. |
|
Jacket | Silicone Rubber |
| Dia. of round assembly under jacket (inch) | Minimum average thickness (mils) | Minimum thickness at any point (mils) |
| 0.200 or less | 30 | 24 |
| 0.201 - 0.300 | 45 | 36 |
| 0.301 - 0.400 | 60 | 48 |
|
Standard | Appliance Wiring Material UL 758. |
|
Marking | General. |
|
Use | Interconnecting Cable for Medical Instruments. |
|