Rating | 150 deg C, 125 Vac, Cable flame. |
|
Conductor | 36 AWG - 20 AWG solid or stranded round. |
|
Insulation | Extruded PFA, 7 mils min average, 6 mils min at any point. |
|
Assembly | Two or more insulated singles optionally twisted around a fluoropolymer filler or breather tube. Fillers may be used to form round cable. |
|
Inner Jacket | Optional, SR (Silicone rubber), 30 mils min average, 24 mils minimum at any point. |
|
Shield | Optional. |
|
Outer Jacket | SR (Silicone Rubber); |
| thickness of core od in inches | Minimum average thickness | Minimum thickness at any piont |
| 0.200 or less | 30 mils | 24 mils min at any point |
| 0.201-0.300 | 45 avg | 36 min |
| 0.301-0.400 | 60 avg | 48 min |
|
Standard | Appliance Wiring Material UL 758. |
|
Marking | General. |
|
Use | Interconnecting Cable for Medical Instruments |
|