Rating | 60 deg C, 125 Vac, Horizontal flame. |
|
Conductor | 30 AWG - 20 AWG, stranded. May include mixed AWG sizes. |
|
Insulation | Extruded PVC, 12 mils minimum average thickness, 10 mils minimum thickness at any point. |
|
Assembly | Two to eight conductors wound in retractile form. |
|
Jacket | Extruded PVC, 60 mils minimum average thickness, 48 mils minimum thickness at any point. |
|
Standard | Appliance Wiring Material UL 758. |
|
Marking | General. |
|
Use | Internal ac interconnecting Cable for Medical and Dental Equipment. |
|