Rating | 60 deg C, voltage not specified, Horizontal flame. |
|
Conductor | For Control conductor - 24 AWG, stranded, minimum number of strands, 37.
For other conductor - 18 AWG, stranded, stranding same as for Type SJT cord. |
|
Insulation | For Control conductor - extruded FEP, Style 1226.
For other conductor - extruded EP, 50 mils min average, 45 mils minimum at any point. |
|
Conductor Assembly | A - Four Control conductors cabled around other conductor. Fillers as needed to make round.
B - Same as A without Control conductors. |
|
Jacket | Extruded Neoprene, 50 mils min average, 40 mils minimum at any point. Extruded over insulated conductor. |
|
Cable Assembly | Two components laid parallel. Components (shown in Conductor Assembly) may be either one of A and one B, or two of A. |
|
Overall Jacket | Extruded Neoprene, 50 mils min average, 40 mils minimum at any point. |
|
Standard | Appliance Wiring Material UL 758. |
|
Marking | General. |
|
Use | Internal defibrillator paddle cable where the acceptability of the combination has been determined by Underwriters Laboratories Inc. |
|